Service manual
GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 64
Trigger Window (0018,1094) 3 Sent if available, else not sent
Reconstruction Diameter (0018,1100) 3 Always sent. (even if value is zero length)
Receiving Coil (0018,1250) 3 Sent if available, else not sent.
Transmitting Coil (0018,1251) 3 Sent if available, else not sent.
Acquisition Matrix (0018,1310) 3 Sent if available, else not sent.
Phase Encoding Direction (0018,1312) 3 Sent if available, else not sent. Is ‘ROW’
or ‘COL ‘ if sent.
Flip Angle (0018,1314) 3 Sent if available, else not sent.
Variable Flip Angle Flag (0018,1315) 3 Always sent. Value is ‘Y’ for
RAMPED_x: where x is AP, IS, IR, PA,
SI, or AL, else value is ‘N’.
SAR (0018,1316) 3 Sent if available, else not sent.
Temporal Resolution (0020,0110) 3 Always sent. (even if value is zero length)
A.3
SC Specific IOD Definition
A.3.1
SC Image IOD Modules
Entity Name Module Name Reference Usage
Patient Patient A.5.1 M
Study General Study A.5.2 M
Patient Study A.5.3 U
Series General Series A.5.4 M
Equipment General Equipment A.5.5 U
SC Equipment A.5.13 M
Image General Image A.5.6 M
Image Pixel Only the type 1 elements are sent. M
SC Image Not sent (consists entirely of type 3 element) M
Modality LUT Only the type 1C elements are sent. U
VOI LUT A.5.11 U
SOP Common A.5.9 M
A.4
Standalone Overlay Specific IOD Definition
A.4.1
Standalone Overlay Image IOD Modules
Entity Name Module Name Reference Usage
Patient Patient A.5.1 M
Study General Study A.5.2 M
Patient Study A.5.3 U
Series General Series A.5.4 M
Equipment General Equipment A.5.5 M
Overlay Overlay Identification A.5.14 M
Overlay Plane A.5.10 M
SOP Common A.5.9 M
A.5
CT/MR/SC/Standalone Overlay IOD Common Table Definitions
A.5.1
Patient Module
Attribute Name Tag Type Notes
Patient’s Name (0010,0010) 2 As entered at the user interface or from
worklist. Supports 5 different components