Service manual
GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 59
0002 )
>Photometric Interpretation ( 0028,
0004 )
MONOCHROME2 MONOCHROME2
>Rows ( 0028,
0010 )
Image Dependent Image Dependent
>Columns ( 0028,
0011 )
Image Dependent Image Dependent
>Pixel Aspect Ratio ( 0028,
0034 )
1/1 1/1
>Bits Allocated ( 0028,
0100 )
8 8
>Bits Stored ( 0028,
0101 )
8 8
>High Bit ( 0028,
0102 )
7 7
>Pixel Representation ( 0028,
0103 )
0(unsigned integer) 0(unsigned integer)
Polarity ( 0020,
0020 )
NORMAL NORMAL
If Failure status is returned during N-SET operation of Image Box following action will
be taken by Print SCU.
0xC605 “Resources temporarily not available” message will be sent.
All other status “Failure” message will be sent.
In all the cases the print session will be terminated.
4.2.3.1.3.4 Printer SOP Class
N-GET DIMSE service is supported for Printer SOP Class. If an N-EVENT-REPORT
DIMSE service is received when association is active, Print SCU handles the relevant
states but the data received is ignored.
The Print SCU issues the request to retrieve following attributes.
If Printer returns Printer Status of FAILURE the association is terminated and the requester
will be notified of the printer status. If the WARNING status is received the SCU will
continue with the print session.
Optional Attribute DICOM Tag Default Value
*Printer Status ( 2110, 0010 ) Printer shall return Value
*Printer Status Info ( 2110, 0020 ) Printer shall return Value
*Printer Name ( 2110, 0030 ) Printer shall return Value
*Manufacturer ( 0008, 0070 ) Printer shall return Value
*Manufacturer Model
Name
( 0008, 1090 ) Printer shall return Value
*Device Serial No. ( 1800, 1000 ) Printer shall return Value