Service manual
GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 58
REPLICATE
NONE
Max Density ( 2010,
0130 )
0-4095 Set in Configuration File
Configuration
Information
( 2010,
0150 )
Printer Dependent Set in Configuration File
Smoothing type ( 2010,
0080 )
Printer Dependent Set in Configuration File
Border density ( 2010,
0100 )
BLACK
WHITE
Set in Configuration File
Empty image density ( 2010,
0110 )
BLACK
WHITE
Set in Configuration File
Min density ( 2010,
0120 )
0-4095 Set in Configuration File
Note: Attributes “sent zero length” use the camera default values.
If Failure status is returned during N-CREATE operation of Film box following action will
be taken by Print SCU.
0x213 “Resource Limitation” message will be sent.
0x106 “Unsupported Film Format” message will be sent.
All other status “Failure” message will be sent
In all the cases the print session will be terminated.
If Failure status is returned during N-ACTION operation of Film box following action will
be taken by Print SCU.
0xC602 “Unable to Create Print Job” message will be sent.
All other status “Failure” message will be sent.
In all the cases the print session will be terminated.
If Failure status is returned during N-DELETE operation of Film box following action will
be taken by Print SCU.
All the return status “Failure” message will be sent and the print session will be
terminated.
4.2.3.1.3.3 Basic Grayscale Image Box SOP Class
The Print SCU supports the following DIMSE Service Elements for Grayscale Image Box
SOP Class.
N-SET - Requests the Printer to set the image box attributes.
The Following Attribute values are supported.
Attribute DICOM Tag Valid Range Default Value
Image Position ( 2020,
0010 )
Based on Image
Display format
No Default Value
Preformatted Grayscale
Image Sequence
( 2020,
0110 )
>Samples per pixel ( 0028, 1 1