4.2.2 Association Establishment Policy
4.2.2.1 General
The Print SCU provides options to indicate the printer AE title, Printer host name. In this
product these inputs are provided by Film Composer. Depending on these inputs the Print
SCU will establish the association with the desired printer.
The DICOM Application Context Name (ACN), which is always proposed, is:
Application Context Name 1.2.840.10008.3.1.1.1
The Maximum Length PDU negotiation is included in all association establishment
requests. The maximum length PDU for association initiated by the DICOM Print SCU AE
is:
Maximum Length PDU 10 Kbytes
4.2.2.2 Number of Associations
The Print SCU will initiate only one association with printer. This will not initiate any other
associations while the current association is active.
4.2.2.3 Asynchronous Nature
The print SCU does not support asynchronous operations. All operations will be performed
synchronously.
4.2.2.4 Implementation identifying information
The Implementation UID allows unique identification of a set of products that share the
same implementation.
The Implementation UID for this GEMS Implementation is:
SIGNA PROFILE Implementation UID 1.2.840.113619.6.93
4.2.3 Association Establishment Policy