Service manual
GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 47
The modality attribute (0008,0060) in the DICOMDIR Series records should be CT/MR for Seconday
Capture and Standalone Overlay SOP Class Images. Otherwise the SIGNA PROFILE DICOM media
display browser will not list-up the series for contents rendering the retrieval to local storage impossible.
3.3.3.1.4.1 Additional Keys
None.
3.3.3.1.5 Other Parameters
This section defines other parameters common to all specific Application Profiles in the STD-CTMR-
MOD12 class which need to be specified in order to ensure interoperable information interchange.
3.3.3.1.5.1 Image Attribute Values
The attributes listed in Table 3.3.3.1.5.1-1 used within the CT/MR/SC/Standalone Overlay Image files, shall
take the values specified.
Table 3.3.3.1.5.1-1: CTMR-MOD Required Image Attribute Values for CT and MR, SC and OV
Images
Attribute Tag Value
Modality (0008,0060) CT/MR
Photometric Interpretation (0028,0004) MONOCHROME2
Bits Allocated (0028,0100) 16
Bits Stored (0028,0101) 16
Overlay data if present shall be encoded in Overlay Data (60XX,3000).
3.3.3.1.5.1.1 Attribute Value Precedence
None.
3.3.3.2 STD-CTMR-MOD23 Class Profile
3.3.3.2.1 SOP Classes and Transfer Syntaxes
This class of Application Profiles is based on the Media Storage Service Class with the Interchange Option.
(see PS3.4).
SOP Classes and corresponding Transfer Syntaxes supported by the STD-CTMR-MOD23 Application
Profiles is also specified in the Table 3.3.3.1.1-1.
3.3.3.2.2 Physical Media and Media Formats