Service manual

GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 39
3 MEDIA STORAGE CONFORMANCE STATEMENT
3.0 Introduction
This Conformance Statement (CS) specifies the SIGNA PROFILE compliance to DICOM v3.0
Media Interchange. It details the DICOM Media Storage Application Profiles and roles which are
supported by this product in its version 7.71.
Note the format of this section follows the format of the DICOM Standard Part 2 (conformance)
Annex C hence the paragraph numbering scheme. Please refer to that part of the standard while
reading this section.
3.1 Implementation Model : MOD Archive Server
All DICOM functionality on the SIGNA PROFILE product is handled by the DICOM Server Application
Entity (AE). The DICOM Server AE is commanded to perform DICOM services through the buttons and
menu selections on the main user interface panel.
3.1.1 Application Data Flow Diagram
The MOD Archive Server creates and/or updates the 5.25 inch (130mm) MOD media with various DICOM
SOP instances. It can process CT, MR, Secondary Capture and Overlay IOD's.
Illustration 3–1 Specific AE Application Model
DICOM MOD
Archive Server
AE
Display
Directory
FSC
FSR
Create
MOD
FSU
Media
Storage
Local
Storage
Update
MOD
The MOD Archive Server AE has a local storage that may contain various SOP instances. These may have
been obtained by original creation, network (DICOM or proprietary) or by removable media using other
application entities. These instances are external to this conformance claim and the origin of SOP instances
is outside the scope of this claim.