GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 38
2.4.4 Private Data Elements
Refer to Appendix B for a complete listing of private data elements used with this implementation.
2.5 CONFIGURATION
2.5.1 AE Title/Presentation Address Mapping
The SIGNA PROFILE system allows the user to add, Remove, or Update the
mapping of remote AE Titles to IP Addresses and Ports. These options can be
selected from the Remote Host Selection menu displayed by choosing Selected
remote host: from the Network pull-down menu from the local database
manager.
2.5.2 Configurable Parameters
The following fields are configurable for the DICOM Server AE:
Local AE Title (the machine hostname)
Local IP Address
Local IP Netmask
Max PDU length
Time-outs,which are set for all hosts, are configurable in dcs.cfg:
note: * denotes any SOP class (time in sec)
*Association time-out - bi_assoc_tio
*Inactivity time-out - bi_store_tio
*Session timeout - bi_session_tio
*Move operation time-out - bi_move_tio
*Find time-out - bi_find_tio
Note: All configurations should be performed by a GE Field Service Engineer.
Note: The local port on which the SIGNA PROFILE system receives DICOM incoming
TCP connections is port 4006.
2.6 SUPPORT OF EXTENDED CHARACTER SETS
In addition to the DICOM default character set, SIGNA PROFILE supports the
ISO IR 100 Latin alphabet #1 supplementary set for the purpose of
interchange.