Service manual
GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 23
Description Type Tag Value
Study date R 0008,0020 Zero length for any Study date,
or “yyyymmdd-yyyymmdd”
matches inclusive range of
‘from-to’ Study dates. ‘From’ or
‘To’ date can be zero length
meaning that side of the range
is any date.
Study time R 0008,0030 Zero length
Accession number R 0008,0050 Zero length for any Accession
number, or Accession number
matches this value
Patient’s name R 0010,0010 Zero length for any Patient
name, or Patient’s name
matches this value
Study id R 0020,0010 Zero length for any Study id, or
Study id matches this value
Study Instance UID U 0020,000D Zero length for Study level
query. Study Instance UID of
study for which matches are
requested at a lower level
Study description O 0008,1030 Zero length
Private Creator Identification P 0009,0010 GEMS_IDEN_01
Suite Id P 0009,1002 Zero Length
Table 2.2.1.2.2.2.1-2: Requested Series Level Keys
Description Type Tag Value
Modality R 0008,0060 Zero length
Series number R 0020,0011 Zero length
Series Instance UID U 0020,000E Zero length for Series
level query. Series
instance UID of series
for which matches are
requested at a low
level
Series description O 0008,103E Zero length
Manufacturer O 0008,0070 Zero length
Images in series O 0020,1002 Zero length
Table 2.2.1.2.2.2.1-3: Requested Image Level Keys
Description Type Tag Value
Image number R 0020,0013 Zero length
Image Instance UID U 0008,0018 Zero length
Image type O 0008,0008 Zero length
Rows O 0028,0010 Zero length
Columns O 0028,0011 Zero length
Image position O 0020,0032 Zero length
Image orientation O 0020,0037 Zero length
Slice thickness O 0018,0050 Zero length
Slice spacing O 0018,0088 Zero length
Gantry tilt O 0018,1120 Zero length