Service manual
GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 14
2 SECTION 2 - NETWORK CONFORMANCE STATEMENT
2.0 INTRODUCTION
This Conformance Statement (CS) specifies the SIGNA PROFILE compliance to DICOM
v3.0. It details the DICOM Service Classes and roles which are supported by this product
in it’s version 7.71.
The SIGNA PROFILE product uses DICOM services to import images for possible further
analysis and/or processing. It also uses DICOM services to export images to other
DICOM-compliant machines.
Note the format of this section follows the format of the DICOM Standard Part 2
(conformance) Annex A hence the paragraph numbering scheme. Please refer to that part
of the standard while reading this section.
2.1 IMPLEMENTATION MODEL
All DICOM functionality on the SIGNA PROFILE product is handled by the DICOM
Server Application Entity (AE). The DICOM Server AE is commanded to perform
DICOM services through the buttons and menu selections on the main user interface
panel. The DICOM Server AE is also listening to a pre-defined port for incoming
connections.