User Manual
7
www.cbsscientic.com
SECTION 2
Description of Parts
2.1 Unpacking
PleaseverifythatyourMidi-HorizontalKitcomescompletewiththe
followingcomponents:
2.2 Components/Assembly
SECTION 3
Instructions for Use
3.1 Midi-Gel Unit Preparation and Gel Casting
1.PlaceMidi-Horizontalunitonalevelwork
surfaceinanauthorizedworkarea.Place
theadjustableheightcombassemblyover
thegelbedorgeltray.Loosenthewhite
screwsholdingthecombtothebacking
andadjustcombdepth(generally1mm).
Gentlytightentheadjustingscrewsand
setthecombaside.SeeTablesonpages
11-12forcombvolumes.
2.Powerleadsareshipped
separatelyandmust be at-
tached before using your
Midi-Gel. Threadthepower
leadendsintothelidrecep-
taclesandrotateinclockwise
directiontohandtighten.
3.2 Stain Options
1) Safe Stains
SafeStainsarealessmutatgenicalternativetoDNAandRNAvisualizationthan
EthidiumBromideanddonotrequireexposuretoUVlight.FollowtheSafeStain
manufacturer’sinstructionsandrecommendations.Safestainscanbeaddedat4
differentstagesofelectrophoresis:
a)intotheagarosegelsolution-Sections3.2.5and3.3.2
b)intorunningbuffer-Section3.4.1
c)intosampleloadingbuffer-Section3.4.2
d)afterelectrophoresis-Section3.5.2
2) Ethidium Bromide (EtBr) Staining
EtBrisaMutagen.ALWAYSweargloves.Disposeofsolutionsinaccordancewith
thesafetyregulationsofyourinstitution.EtBrcanbeaddedat3differentstagesof
electrophoresis:
a)intotheagarosegelsolutionafterthesolutionhasbeenmademoltenand
thetemperaturehasdroppedto55°C-Sections3.2.5and3.3.2
b)intorunningbuffer-Section3.4.1
c)aftertheelectrophoresisruninastainsolution.Useatanalconcentration
of0.5-1µg/ml.-Section3.5.2