SECTION 1
General Information
1.1 Introduction
Electroblottingisafaster,easier,andmoreefcientmethodtopermanentlytransferPAGEgelstonitrocellulose
andnucleicacidstonylon,thancapillary(passivediffusion)blotting.Forasimilartransfer,traditional“capillary
blotting”cantake6-24hourstocompletecomparedtoonly2hoursorlessusingelectroblotting.TheC.B.S.Sci-
enticElectrophoreticBlottingSystemscomeequippedwith;twoelectrodepanels,onecolor-codedgelcassette
assembly,electrophoresistankwithcoolingchamberandsafetycoverwithattachedpowerleads.TheEBU-222
gelcassettesizeis20cmx21.5cmandcantransferuptothreegelsatonetime.TheEBU-222gelcassettesize
is9x11cmandcantransferupto2gelsatonetime.
1.2 Specications
Constructions:
Bufferchamber,safetycover Acrylic
Electrodepanels Acrylicandplatinumwire.010”diameter
Powercords FRPolypropylene/Silicone,rated7500V,200mA,65° C
Gel cassette Perforated acrylic panels red and black, two
sponge pads
SafetyCertication EN61010-1-1993(IEC1010-1)
Shipping Weight 5 lbs.
OverallSize(l)x(w)x(h)cm 19x11x28.5
MaximumGelSize-cm 18x18
Distancebetweenelectrodes-cm 8.7
Recommendedbuffervolume
1 cassette 3.5 liters
2 cassettes 3.0 liters
1.3 Safety
PowertotheElectro-BlotteristobesuppliedbyanexternalDCvoltagepowersupplythatmustbeground
isolatedsothattheDCvoltageoutputoatswithrespecttoground.Foranypowersupplyused,themaximum
speciedoperatingparametersfortheunitsare:
MaximumLimits
100 VDC
70 watts power
1500mAcurrent
50ºCambienttemperature
Currenttotheunit,providedfromtheexternalpowersupply,mustentertheunitthroughthelidassembly,provid-
ingasafetyinterlocktotheuser.Currenttotheunitisbrokenwhenthelidisremoved.Donotattempttousethe
unitwithoutthesafetylid,andalwaysturnthepowersupplyoffbeforeremovingthelid,orwhenworkingwiththe
unitinanyway.Followsafetyprecautionsspeciedbythepowersupplymanufacturer.