3.6GelCastingTechniques-continued
B.VerticalgradientgelcastingusingGM-40gradientmakerandgravityowforDGGE.
1. Place the GM-40, gradient maker, on an elevated magnetic stirrer with a small “ea” stir bar in
the cylinder (C-2) closest to the outlet. The gradient maker should be tted with a luer valve, (V-
2) and a 20ga needle with attached tubing to deliver acrylamide to the gel plate sandwich. (Fig
3-4). With the height differential between the luer valve (V-2) and the top of the gel plate sandwich
between 2-4 inches, it takes approximately 5-9 minutes to cast the gel. Raising the height of the
gradient maker can increase ow rate. Flow can be regulated by turning luer valve (V-2). Flow
rate appropriate for each gradient should be determined empirically.
C-1
C-2
V-1
V-2
Fig. 3-4
2. For TTGE and SSCP single percentage gels, use a pipette or syringe to inject polyacrylamide into
glass plate cassette.
 Refertosection3.7forGradientGelFormulaProtocol.
See section 4.3 for running conditions.