Manual

12
www.cbsscientic.com
3.2 Heater/Stirrer Set-Up and Programming – continued
3.2cSpecications
Temperature Stability
1
±0.01°C
Controller / RS232 Yes
External Temperature Probe Functional on Programmable models / optional external
probe required
Not functional on Digital models
Readout Accuracy Graphic LCD, °C or °F, ±0.25°C
Heater 1100W – 115V, 2200W – 240V
Pressure Flow Rate 15.7 to 30 LPM (60Hz); 15 to 22 LPM (50Hz)
Suction Flow Rate 11 to 22 (60Hz); 10 to 15 LPM (50Hz)
Over-Temperature Protection Yes, user-adjustable
Low-Liquid Protection Yes
Pump Speeds Five-speed adjustable
1. Temperature stability may vary depending on bath volume, surface area, insulation, and type of uid.
NOTE: Performance specications determined at ambient temperature of 20°C (68°F).
Environmental Conditions:
• Indoor Use Only • Over Voltage: Category ll
• Maximum Altitude: 2000 meters • Operating Ambient: 5° to 30°C
• Relative Humidity: 80% for temperatures to 30°C • Pollution Degree: 2
• Class 1: Residential, Commercial, Light Industrial • Class 2: Heavy Industrial