User Manual

C.B.S. Scientific 25 Dedicated Height Sequencer
5.3 Comb/Well Volumes
Sequencing Combs for 20cm wide units
Note: To calculate sample well volume expressed in millimeters (mm) divide maximum volume by tooth depth.
-Volume per tooth (microliters-l)
# of wells Tooth
width/mm
0.2mm thickness 0.4mm thickness 0.5mm thickness
8 17.7mm
33.6l 67l 84l
12 11mm
21l 41l 52l
14 9mm
17.3l 34l 43l
16 7.6mm
14.4l 28l 36l
17* 6.6mm
12.5l 25l 31.3l
20 5.6mm
10.6l 21l 26.5l
28 4mm
7.8l 15l 19.5l
32 3.4mm
6.4l 12l 16.5l
35* 2.92mm
5.5l 11.12l 13.9l
*Designed for multi-channel pipettor loading
Overall length of comb: 15.87cm
Tooth depth: 9.5mm
Spacing between teeth: 8-20 wells : 2.3mm
28-35 wells: 1.5mm
Sequencing Combs for 33cm wide units
Note: To calculate sample well volume expressed in millimeters (mm) divide maximum volume by tooth depth.
-Volume per tooth (microliters-l)
# of wells Tooth
width/mm
0.2mm thickness 0.4mm thickness 0.5mm thickness
8 33.4mm
63.4l 127l 158.6l
16 15.5mm
29.5l 59l 73.6l
24 9.5mm
18l 36l 45l
31* 6.6mm
12.5l 25l 31.4l
34 6mm
11.4l 22l 28.5l
50 4.1mm
7.7l 15.5l 19.4l
60 3.1mm
5.8l 11.6l 14.7l
63* 2.92mm
5.5l 11.1l 13.8l
*Designed for multi-channel pipettor
Overall length of comb: 28.4cm
Tooth depth: 9.5mm
Spacing between teeth: 8-31 wells : 2.3mm
34-63 wells: 1.5mm
Ultrathin Sharktooth Combs for 20cm wide units
Note: To calculate sample well volume expressed in millimeters (mm) divide maximum volume by tooth depth.
-Volume per tooth (microliters-l)
# of wells Tooth width/mm .075mm thickness
1 158.7mm
47.6l
24 5.92mm
1.6l
34* 4.5mm
1.2l
38 4.02mm
1.0l
52* 3.0mm
.8l
*Designed for multi-channel pipettor Point to Point Measurements:
24 well=5.92mm
34 well=4.50mm
38 well=4.02mm
52 well=3.00mm
Ultrathin Sharktooth Combs for 33cm wide units
Note: To calculate sample well volume expressed in millimeters (mm) divide maximum volume by tooth depth.
-Volume per tooth (microliters-l)
# of wells Tooth width/mm .075mm thickness
1 284mm
85.20l
46 5.92mm
1.6l
62* 4.5mm
1.2l
68 4.02mm
1.0l
94* 3.0mm
.8l
*Designed for multi-channel pipettor Point to Point Measurements: