User Manual

8
Wide Dual Mini-Vertical Instructions 9-30-13
SECTION 3
Instructions for Slab Gel Electrophoresis Using Pre-Cast Gels
3.1 Preparing the Electrophoresis Unit
1. Power leads are shipped separately and
must be attached before using your
Wide Dual Mini-Vertical. Thread the
power lead ends into the lid receptacles
and rotate in clockwise direction to hand
tighten.
2. Place unit in authorized work area. Re-
move safety cover from the assembled
unit by simultaneously pressing down on
white push pins while lifting up on blue
safety cover as shown in gure 1.
Do not remove safety cover by pulling
up on leads!
3. Remove white core from lower reservoir
by grasping core with one hand and lift-
ing directly up as shown in gure 2.
4. Open doors on the core assembly by
pulling up on the white latches, as shown
in gure 3.
1
2
3