User guide

11
www.cbsscientic.com
matching
colors
8
9
4. Attach safety cover and turn on magnetic
stirrer. The closed unit ready for power is
shown in gure 8.
5. Connect the leads to the power supply,
matching the color-coded red to red and
black to black. See Section 6.1 for rec-
ommended power conditions. Begin
separation by electrophoresis.
3.3 Removing the Gel
1. Turn the power supply off and discon-
nect the leads from the power supply.
Remove the safety cover from the unit,
by placing thumbs on white posts next
to red & black connectors, then pushing
down while pulling up with ngers under
lid as shown in gure 9.
Do not remove safety cover by pulling
up on leads!
2. Pull up on gel door latches, and open gel
door. Remove gel sandwich from Cas-
sette Assembly. Stain and x according
to your preferred method.