Section Section 5 Maintenance of Equipment
5.1 Care and Handling
ThecomponentsoftheMega-Gelarefabricatedfromacrylic,Delrin®,Polycarbonate,glassandanodizedaluminum.
Aswithanylaboratoryinstrument,adequatecareensuresconsistentandreliableperformance.
Aftereachuse,rinsebufferreservoirs,combs,spacers,clampsandglassplateswithdeionizedwater.Wipedry
withasoftclothorpapertowel,orallowtoairdry.Whenevernecessary,allcomponentsmaybewashedgentlywith
waterandanonabrasivedetergent,andrinsedanddriedasabove.Neveruseabrasivecleaners,windowsprays
orscouringpadstocleanthecomponents,asthesecausedamagetotheplasticsurfaces.
•Donotautoclaveordry-heatsterilizetheapparatusorcomponents.
•Donotexposetheapparatusorcomponentstophenol,acetone,benzene,halogenatedhydrocarbon
solventsoralcohols.
•AvoidprolongedexposureoftheapparatusorcomponentstoUVlight.
5.2 Maintenance
The following simple, routine inspection and maintenance procedures will help ensure both the safety and the
performanceoftheMega-Gel.Fororderinginformationofreplacementpartscall1-858-755-4959.
•Inspectelectricalconnectionsandpowercordsregularly.Ifthebananaplugconnectorstoeithertheunitorthe
powersupplydonotexhibitreasonablefrictionorcanberockedeasily,replacetheplugsorpowercords.
•Ifpowercordsshowanysignsofwearordamage(e.g.cracks,nicks,abrasions,ormeltedinsulation),replace
themimmediately.
•Examinethebananajackstoensurethattheyaretightandfreeofcorrosion.Ifbananajacksarecorroded,
theyshouldbereplaced.