User Manual
11 Mega-Gel Instructions 12/11
9. ConnecttheDCpowerleadstothepowersupplywiththeproperpolarity.Makesuretheblackleadsareconnected
totheblackcathode(-)andtheredleadsareconnectedtotheredanode(+).
10.Rinsesamplewellswithupperreservoirbuffer.Beforeloadingsamples,pre-electrophoresethegelfor20minutes.
WARNING: Excessivepowerwillcausethegeltooverheatandcracktheglassplates.
4.2 Loading Samples
1. Attheendofthepre-electrophoresisperiod.TURN OFF THE POWER SUPPLY AND DISCONNECT BOTH
DC POWER CORDS FROM THE POWER SUPPLY.Openupperreservoirsafetycover.
2. Immediately prior to loading the samples, rinse the wells of the gel thoroughly with electrophoresis buffer.
Use a loading tip or a syringe with a bent needle to wash away urea that has diffused into the wells.
4.3 Comb Volume/Maximum well volume
Note:Tocalculatesamplewellvolumeexpressedinmillimeters(mm)dividemaximumvolumebytoothdepth.
-Maximum Volumepertooth(microliters-ml)
#ofwells Tooth
width
0.75mm
thickness
1.0mm
thickness
1.5mm
thickness
100 3mm 22.5ml 30ml 45ml
Overalllengthofcomb:44.8cm
Toothdepth:10mm
Spacingbetweenteeth:1.5mm
4.4 Electrophoresis
1. Afterloadingthesamples,attachthereservoirsafetycover.ConnecttheDCpowerleadstotheunitthentothe
powersupplywiththeproperpolarity.Makesuretheblackleadsareconnectedtotheblackcathodes(-)and
theredleadsareconnectedtotheredanode(+).
2. Turnonthepowersupplyandsetthevoltageorwattagetothepropersettingforthegel. 350V at 105mA/ gel
at 0.5X TBE buffer for 1.5mm thick gel.
3. Monitortheprogressofelectrophoresisbyfollowingmigrationofamarkerdyefrontorbyanyotherpreferred
method.
4. Whenelectrophoresisiscomplete,TURN POWER SUPPLY OFF.Disconnectall DCpowercords,rstfrom
thepowersupply.