User Manual
10Mega-Gel Instructions 12/11
Section 4 Electrophoresis
4.1 Pre-electrophoresis
1. Filllowerbufferreservoirwithbuffer.
2. After removing the Gel Wrap Gasket from the gel sandwich, but with the white clamps still in place,
rinse the sandwich in D.I. water and dry. Remove comb, then remove the white plastic clamps and
place sandwich (notched plate inward) into the “buffer- lled “ lower resevoir at a slight angle to prevent
capturing bubbles. Reclamp sandwich to unit using bar clamp as quickly as possible as described below.
3. Pre-loosenblackwingnutsonbarclampsoclampassemblyslidesoverglassplatesandupperreservoir.
Tightenwingnutsonlyenoughtocompressgasket.DoNOTovertighten.Seeguresbelow.
4. Liftoutbasesecuringpinandrotatebase180ºforeasyaccesstooppositeside.Repeatsteps1-3withsecond
setofglassplates.
5. Filleachupperreservoirwithapproximately500mlofrunningbuffer.Makesurenobufferisleakingfromtheupper
bufferreservoir.Filleachlowerreservoirwithapproximately500mlofelectrophoresisbuffer.Besurenobubbles
obstructbuffercontactwiththeloweredgeofthegel.Note:DoNOToverllbufferreservoirsorallowbufferto
makecontactwiththebananajacks.
WARNING: No buffer should leak through or around the silicone gasket or down the side of
the gel assembly. Leakage may allow the upper reservoir to drain dry or cause
arcing and damage to the apparatus.
6. Attachthereservoirsafetycoverstotheupperandlowerreservoirs.Connectthefemaleendoftheblackpower
cordintothesafetycoveroftheupperreservoir.Connectthefemaleendoftheredpowercordintothesafety
coverofthelowerreservoir.
7. Connectanexternalchillertothetubingadaptersonthesideoftheaparatusmakingsuretousethebottomtubing
adpaterastheinletandthetopastheoutlet.
8. BesuretoattachTemperatureStrip(Cat.#EGT-100)toupperormiddleofglassplate,tomonitorthetemperature
ofthegel.