Maintenance and Troubleshooting
Recommended Maintenance
1� Antenna connections should be visually inspected for corrosion,
contamination or water ingress on an annual basis�
2� Any maintenance required is limited to the external enclosure
surface, power cable connections, and the firmware which can be
updated by using a WiFi capable computer and the embedded web
server used for configuration� Any unit that continues to exhibit a
failure mode after performing the troubleshooting steps outlined
below should be returned to BW Technologies by Honeywell for
maintenance, repair, or replacement� There are no serviceable
components inside the enclosure
Troubleshooting LocaXion Anchor
The LocaXion Anchor does not have any user-serviceable parts� Any
failure within the LocaXion Anchor requires a hardware replacement
There are limited ways to determine whether a LocaXion Anchor has
failed� The LEDs provide a visual indication that the LocaXion Anchor
has power and that the radio module is ready to receive communica-
tions from a client or laptop computer
1� Confirm that the red LED labeled POWER is on� If the red POWER
LED fails to turn on confirm that power is available and all connec-
tions have been made correctly
2� If the red POWER LED is on confirm that the green STATUS LED
turns on within 10 seconds (possibly longer if the unit has been left
unpowered for an extended period of time at temperatures at or
below -20 degrees centigrade) of the POWER LED turning on�
3� In the event that the Green STATUS LED does not turn on the unit
may be recovered by either power-cycling the unit or by bringing a
magnet up to the target area (unused knock-out port to the left of
the LED label) and removing it after 1 second� The magnet must
be suitably strong and be brought in contact with the target area
to ensure that the module is reset� After removing the magnet the
green STATUS LED should come on within 10 seconds thereby
indicating that the LocaXion Anchor has been successfully reset
and should now be fully functional�
4� If the LEDs continue to indicate that a problem exists after confirm-
ing power is available, connections are secure, and/or the module
has gone through the reset process than the unit must be returned
to BW Technologies by Honeywell for further evaluation and repair