User Manual
ImpactXtreme Operator’s Manual 8 March 2013
f)
Vibrator alarm activates.
OL Alarm
Detector Behavior
a)
Gas channel displays alarm icon.
b)
Heartbeat icon disappears.
c)
LCD backlight changes to red.
d)
LED visual alarms flash fast, red.
e)
Audible alarm is fast with downward tone.
f)
Vibrator alarm activates.
Multi-Gas Alarm
Detector Behavior
a)
Gas channel display alternates between alarm icons and gas concentrations.
b)
Heartbeat icon disappears.
c)
LCD backlight changes to red.
d)
LED visual alarms flash fast, red.
e)
Audible alarm alternates between low alarm and high alarm tones.
f)
Vibrator alarm activates.
Battery Alarms
The ImpactXtreme tests battery power during start-up, and continuously throughout operation.
When there is only enough power left to operate in a normal atmosphere for about 15 minutes,
the low battery alarm activates. After about 15 minutes of low-battery alarm, the detector enters
Critical Battery Alarm, and initiates an automatic deactivation procedure. The low battery alarm
is intended to provide enough time to leave a hazardous area before the detector automatically
deactivates.
Low Battery Alarm