User Manual

ImpactXtreme Operator’s Manual 8 March 2013
3)
Begin the bump test.
a) With the detector in normal operating mode, press to open the main menu.
b) Press to scroll to Bump Test, and press to open the bump test menu.
c) Select Normal Bump Test and press to begin the bump test.
4) Select the sensor channels.
a) Press to select or deselect the channels that you will test.
b) Press to move to the next channel.
c) When ready, press to continue to the next step.
5)
When prompted to Apply bump test gas now, connect the detector to the test gas cylinder.
For more information, see Connect/Disconnect Gas Cylinder.
a) Ensure that the calibration cap is fastened securely before applying gas.
6) During the bump test, gas readings fluctuate. The progress bar in the lower-right corner of
the LCD displays bump test progress.
7) When the bump test is complete, the results are displayed.