User's Manual

ConneX1 Operator’s Manual
14
General User Options
A value must be entered in the  field in the Sensor Configuration
section before enabling Force Calibration.
CAUTION

Calibration option is automatically disabled.
The detector is shipped with the Force Calibration option disabled.
Language: Choose the display language. All screens are displayed in that
language. Currently, only English is available. Later releases of firmware will have
other languages.
Locations: List of locations that can be assigned as the default location. Up to
20 locations can be customized.
   This field defines how often the detector records a
datalog (1 to 600 seconds).
The total number of 8-hour days datalogs that can be recorded is assuming 90%
of the day has no gas concentrations.
Table 5. Datalog Storage Capacity
Datalog Interval Total Number of Days Datalogs Can Be Recorded
 
 
 
CAUTION
When the memory is full, the detector replaces the oldest datalogs
with the most recent datalogs.
The detector is shipped with the default setting of 5 seconds.
Operators Name: List of names that can be assigned as the default operator. Up
to 10 operator names can be customized.
Owner Name: Name of the detector owner. Field can be used to identify personnel
responsible for detector upkeep.
 Choosing a profile pre-selects user options appropriate to
the working condition. Refer to Fleet Manager II Operator’s Manual to set up profiles.
Refer to
Appendix A Regional Setpoints on page 56.
Serial Number: Displays the unique serial number of the detector. This is a read-
only field. Serial number is factory defined
Time Format: Select the 12-hour clock or 24-hour clock in time stamps. Select
from 12 (12-hour clock) or 24 (24-hour clock). Default is 12-hour clock.
Wireless Mode Selection: If enabled, the user can enable or disable WiFi. The
detector is shipped with Wireless Mode Selection enabled.
4.1.1 
 Create messages to respond to received messages
sent from LocaXion Manager (20 characters maximum). These messages are
used to communicate with LocaXion Manager.
Enable Startup Message: When checked, a message will display on the detector
LCD during startup.