User's Manual
ConneX1 Operator’s Manual
14
General User Options
A value must be entered in the field in the Sensor Configuration
section before enabling Force Calibration.
CAUTION
Calibration option is automatically disabled.
The detector is shipped with the Force Calibration option disabled.
Language: Choose the display language. All screens are displayed in that
language. Currently, only English is available. Later releases of firmware will have
other languages.
Locations: List of locations that can be assigned as the default location. Up to
20 locations can be customized.
This field defines how often the detector records a
datalog (1 to 600 seconds).
The total number of 8-hour days datalogs that can be recorded is assuming 90%
of the day has no gas concentrations.
Table 5. Datalog Storage Capacity
Datalog Interval Total Number of Days Datalogs Can Be Recorded
CAUTION
When the memory is full, the detector replaces the oldest datalogs
with the most recent datalogs.
The detector is shipped with the default setting of 5 seconds.
Operators Name: List of names that can be assigned as the default operator. Up
to 10 operator names can be customized.
Owner Name: Name of the detector owner. Field can be used to identify personnel
responsible for detector upkeep.
Choosing a profile pre-selects user options appropriate to
the working condition. Refer to Fleet Manager II Operator’s Manual to set up profiles.
Refer to
Appendix A Regional Setpoints on page 56.
Serial Number: Displays the unique serial number of the detector. This is a read-
only field. Serial number is factory defined
Time Format: Select the 12-hour clock or 24-hour clock in time stamps. Select
from 12 (12-hour clock) or 24 (24-hour clock). Default is 12-hour clock.
Wireless Mode Selection: If enabled, the user can enable or disable WiFi. The
detector is shipped with Wireless Mode Selection enabled.
4.1.1
Create messages to respond to received messages
sent from LocaXion Manager (20 characters maximum). These messages are
used to communicate with LocaXion Manager.
Enable Startup Message: When checked, a message will display on the detector
LCD during startup.