User's Manual
ConneX1 Operator’s Manual
13
General User Options
Distributor Name: Name of the distributor company.
Enable Force Bump: If Force Bump is enabled and the sensor is past due for a
bump test, the following screen displays during the startup self-tests.
(Note: “
To start” alternates with the progress bar every 4 seconds.)
The sensor must be bump tested to continue and enter normal operation. Press
, and refer to 9.2 Performing a Manual Bump Test on page 29.
Or
Press and hold
to deactivate the detector.
A value must be entered in the field in the Sensor
Configuration section before enabling Force Bump.
CAUTION
option is automatically disabled.
NOTE
BW Technologies by Honeywell recommends to bump test the
to gas by exposing the detector to a gas concentration that
exceeds the alarm setpoints. Verify that the audible and visual
alarms activate. Calibrate if the readings are not within the
For complete instructions to perform a bump test, refer to
9.2 Performing a Manual
Bump Test on page 29
.
The detector is shipped with the Force Bump option disabled.
Enable Force Calibration: If enabled and a sensor is past due for calibration or
the last calibraiton failed, the following screen displays during the startup self-tests.
The sensor must be calibrated to continue and enter normal operation. Press ,
and refer to
10.4 Manual Calibration Procedure on page 27.
Or
Press and hold
to deactivate the detector.