User's Manual

ConneX1 Operator’s Manual
13
General User Options
Distributor Name: Name of the distributor company.
Enable Force Bump: If Force Bump is enabled and the sensor is past due for a
bump test, the following screen displays during the startup self-tests.
(Note:
To start” alternates with the progress bar every 4 seconds.)
The sensor must be bump tested to continue and enter normal operation. Press
, and refer to 9.2 Performing a Manual Bump Test on page 29.
Or
Press and hold
to deactivate the detector.
A value must be entered in the    field in the Sensor
Configuration section before enabling Force Bump.
CAUTION

option is automatically disabled.
NOTE
BW Technologies by Honeywell recommends to bump test the

to gas by exposing the detector to a gas concentration that
exceeds the alarm setpoints. Verify that the audible and visual
alarms activate. Calibrate if the readings are not within the

For complete instructions to perform a bump test, refer to
9.2 Performing a Manual
Bump Test on page 29
.
The detector is shipped with the Force Bump option disabled.
Enable Force Calibration: If enabled and a sensor is past due for calibration or
the last calibraiton failed, the following screen displays during the startup self-tests.
The sensor must be calibrated to continue and enter normal operation. Press ,
and refer to
10.4 Manual Calibration Procedure on page 27.
Or
Press and hold
to deactivate the detector.