General User Options
1. Activate the detector and wait for the startup sequence to complete.
2. Connect the USB cable to the USB port of the computer.
NOTE
Plug the USB into the same USB port where the USB drivers
were installed.
3. Connect the USB cable to the IR Link.
4. Insert the IR Link into the IR interface on the back of the detector.
5. From the PC, open Fleet Manager II.
The following screen displays when the IR Link is connected to the detector.
6. Click Administration.
7. From the Administration toolbar, click Login / Logout to access the
Enter Password dialog box. Enter Admin and click OK (password is
case sensitive).
8. From the Devices toolbar, click .
9. The Device selection popup displays. Select ConneX1 and click OK.
10. From the configuration window, click Retrieve from Device at the bottom
of the window. The fields automatically populate with the detector’s current
configuration settings.
11. Refer to
4. User Options on page 12 for descriptions about how to enter
data, enable/disable, and define settings.
12. When all settings are defined, click Save to Device at the bottom of the
window. The detector is now updated with the new settings.
4. User Options
User options can be configured using Fleet Manager II.
4.1 General User Options
Define how long (0-30 seconds) the backlight will
stay lit after pressing a button. The default value is 10 seconds. This does not
affect the backlight behaviour during alarms, startup sequence, critical system
faults, and powering off. Default is XX seconds.
If enabled, the sensor can only be calibrated when
calibration is initiated from Fleet Manager II or with the IntelliDoX.
The field defines how often the
confidence beep (5-255 seconds) and confidence flash (1-120 seconds) occur
(The default confidence beep interval is 10 seconds. The default flash interval is
1 second). Enter the desired value.
Date Format: Choose how the date will display, either day/month/year or month/
day/year. Select dd/mm/yyyy or mm/dd/yyyy. Default is mm/dd/yyyy.
Distributor Contact Information: Contact information (phone number or email)
of the distributor.