Product Manual
70-00533-02 A CareCenter MD 1.0
6-14
Acquire a Stress Test ECG
Exercise
Settings
• The protocol options.
• For bicycle stress test:
– The calculation method for the target load (Standard, Jones, Jones
(without usage of weight), or Wassemann).
– Display a message when the destination value is reached.
If the option is cleared, no warning displays. This applies also to the
Load display element described in Work load controls on page 6-20.
– Displays a message containing the load value when the patient reaches
a heart rate of 130, 150, or 170.
You can select pro-rate the load value over the whole stage, otherwise
the system reports the current load.
Note: PWC (physical working capacity) is the load when patient achieves
the heart rate of 130 (PWC130), 150 (PWC150), or 170 (PWC170).
Tab Settings
Configuration Options
Configure the messages • Select File | Options… | 12 Lead ECG | Test procedures and
select the test procedure to change the values.