Stress test displays
Basic settings for limit and target values
You can change the settings on these tabs:
Tab Settings
Heart Rate Setting limits.
To enable limits, select Enable.
Setting the limiting formulas.
To select an absolute value, select Absolute value and enter the value.
Modify the formula for the maximum or the proportional extent of
capacity utilization with geriatric-dependent and gender-dependent
calculation and determine whether or not to use the limit value.
If the Display a message box when the limit is reached option is
cleared, no warning displays. The setting also affects the Heart rate
display element see Display elements on page 6-9.
Modifying the training heart rate formula.
Blood Pressure
The systolic and the diastolic limit values.
If no value is selected, no warning is displayed. This applies also to the
Blood pressure display element described in the Blood pressure on
page 6-9.
The value for the warning at drop of the systolic blood pressure value.
If no value is selected, no warning displays.
If the Display a message box when the limit is reached option is cleared,
no warning displays.
ST settings
The value of the ST elevation and the ST depression and the STmax and
worst case settings.
If no value is selected, no warning displays. This applies also to the
Maximum ST value display element Display elements on page 6-9.
Whether or not the warning should be displayed at ST depression, ST
elevation or with the amount of ST.
If the Display a message box when the limit is reached option is cleared,
no warning displays.
Select which ST measuring point to use.
The channels that are searched for the maximum
The channel selection for Averaged beats.