Product Manual
70-00533-02 A CareCenter MD 1.0
6-12
Acquire a Stress Test ECG
BP decrease > limit Systolic pressure drop too high— compared with
the preceding measurement, the value of the
systolic blood pressure decreased by more than 20
mmHg.
This warning displays only in the exercise phase.
ST depression > limit
ST elevation > limit
ST depression too high—the value of the maximum
ST depression has exceeded the limit value in one of
the leads selected for monitoring.
By default, the maximum ST depression is enabled.
To configure the warnings for the maximum ST
depression, the maximum ST elevation or the
maximum amount of ST for the current test
procedures, select Settings | ST....
The target load value has been reached The target load was reached.
PWC value accomplished at load Indicates the load in the stage when the patient has
achieved a heart rate of 130, 150, or 170.
PWC (physical working capacity) is the load when
patient achieves the heart rate of 130 (PWC130), 150
(PWC150), or 170 (PWC170).
Warning Description