Product Manual

CareCenter MD 1.0 70-00533-02 A
6-11
Stress test displays
Pharmacological Stages
By factory selection, the pharmacological stress test consists of a resting phase and stages 1
through 6 in the exercise phase.
Warnings
During the exercise, the heart rate, the blood pressure values, the ST values, and the load
values are monitored. To be monitored, the parameter must be activated in the test
procedure data. For more information see Stress test parameters on page 6-25.
When an actively monitored parameter reaches the maximum ratings and destination
values, the system indicates warning messages and repeats them every 2 minutes. The text
of the message is entered in the Comments box. You can overwrite the message by
changing the comments.
Note: You must save to see the new comments.
When the warning appears:
To stop the repeated warnings, select Do not warn again in the message.
To save the last 10-seconds of ECG before the warning appeared, click Save 10s ECG.
To close without saving, click Close.
Configuration Options
Enable RPE 1. Select File | Options… | 12 Lead ECG | ECG.
2. In the Stress Settings group, select Use RPE.
Change basic ST values
1. Select: File | Options... | 12 Lead ECG | Test procedures.
2. Select the type of stress test and then select ST Options.
Disable METs
1. Select File | Options… | 12 Lead ECG | ECG.
2. In the Stress options group, clear Use METs.
3. Set the Use actual load or Use target load in the same
group when METs is enabled.
METs with bicycle stress test needs body weight, but with
treadmill it does not.
Warning Description
HR > limit Heart rate too high the value of the heart rate has
exceeded the limit value shown.
BPsys > limit Systolic blood pressure value too highthe value of
the systolic blood pressure has exceeded the limit
value shown.
A corresponding warning displays for the diastolic
blood pressure.