Stress test displays
Averaged Beat panel
The Averaged Beat panel displays the measured ST rate and the ST raise in blue for the
reference beat, and in black for the actual beat. The STmax lead corresponds with that of
the Maximum ST value on page 6-10.
The averaged beat is updated every 10 seconds.
1. Click the lead name to jump to the next lead.
2. To view another lead, right-click on the lead inscription and select the lead.
3. To change the representation, right-click on the averaged beat and select the settings.
ST measuring point
You can measure ST fixed at J+20ms, J+60ms, J+80ms or variable at J+1/8T and J+2/8T.
The variable ST measuring points adapt themselves to the length of the T wave.
◆ Right-click and select Measurement Point, and select the measurement point.
You can toggle the ST measuring point at any time during the test procedure. Also during
later opening and viewing of the record, you can select another ST measuring point.
All analyses, summaries and printouts use the currently selected ST measuring point. See
Basic settings for limit and target values on page 6-13 for how to change the basic setting.