Product Manual
70-00533-02 A CareCenter MD 1.0
6-4
Acquire a Stress Test ECG
The system displays these arrhythmia events:
◆ Asystole
◆ Pause of one complex (1 beat pause)
◆ Pause of two complexes (2 beat pause)
◆ Long RR
◆ Run
◆ Couplet
◆ Ventricular tachycardia
◆ PVC
◆ PAC
◆ Multiple PAC
◆ Bigeminy (Bigeminal pulse)
To... Do this...
Step through multiple
arrhythmia events
• Right-click and select Previous event or Next event.
Exclude or include arrhythmia
events during the stress test
• Select Settings | Arrhythmia detection....
These settings are reset to their original state for the next
stress test.
Configuration Options
Configure which arrhythmia
events to monitor
1. Select File | Options … | 12 Lead ECG | Test procedures.
2. Select the test procedure (for example: Treadmill Test Bruce)
from the list and open the Arrhythmia events tab.
– To exclude an arrhythmia event (the system will not display
the detected event), clear the option.
– To include an arrhythmia event (the system displays the
detected event), select the event.