Product Manual

CareCenter MD 1.0 70-00533-02 A
6-3
Stress test displays
ECG panel
The ECG panel displays the ECG waveforms.
Arrhythmia panel
The Arrhythmia panel displays events that occur during the test. You can disable the
arrhythmia analysis at the beginning of the exercise program. The measurement program
must be running to see arrhythmia events.
The ECG in the Arrhythmia panel is updated every 2.5 seconds. To change speed,
sensitivity, channel position, or elevation of the ECG waveforms see Change the ECG
representation on page 3-19.
The events in the Arrhythmia panel have this information:
The phase in which the event started
The start time of the event
Note: The system displays the phase for the beginning of the event. If the event is long
and uninterrupted such as a bigeminal pulse, the displayed phase and beginning time
will not change, as long as the bigeminal pulse continues. Therefore the bigeminal pulse
can be marked for resting phase, although you have already come to the end of the test
procedure.
The name of the event
Configuration Options
Configure the representation
of the ECG waveforms
1. Select File | Options… | 12 Lead ECG | Test procedures.
2. Select the test procedure (for example: Treadmill Test Bruce)
from the list and open the Display tab.
3. Select the display (for example: Online ECG) and make your
changes.