Product Manual

CareCenter MD 1.0 70-00533-02 A
5-9
Detailed procedures
Review the recording
After the acquisition is complete, the recording displays for confirmation.
If the records are automatically analyzed, the main measurement values, the results of
analysis, the averaged beat and the beat classification are displayed. Records are
automatically measured and analyzed only when:
The test procedure has 12-channel standard leads
The duration of the acquisition is at least 10 seconds
Date of birth and gender of the active patient are first entered or changed
Note: If the test has not been saved, then the system re-analyzes the ECG test whenever age
and gender are changed. For automatic re-analysis:
USBthe acquisition device must remain connected to the system
Wirelessthe acquisition device must remain in acquisition mode
As long as the electrodes are connected to the patient, the running ECG displays in the
rhythm line.
You can view more representations to see more details.
Click the review buttons or select View and select the display.
Rhythm1-channel view of the whole ECG as "a rhythm queue" and multi-
channel ECG strip
ECGmulti-channel ECG of the whole record results of analysis and averaged
beat
ECG Stripmulti-channel ECG, results of analysis and averaged beat of a strip
Measurement Table measurement results of analysis and averaged beat of a strip
Averaged BeatAveraged beat, results of analysis and ECG of a strip
See also Chapter 9, Displaying a Rhythm ECG Record.