Product Manual

CareCenter MD 1.0 70-00533-02 A
5-5
Controls for acquiring a Rhythm ECG
Freeze Frame or F6 Stops the multi-channel ECG, while the rhythm channel and all
measurement functions continue.
1. Click Freeze Frame or press F6.
When the recording is paused, you can turn back the standing ECG
and save ECG strips, mark and print sections, and measure RR
distances.
See ECG view on page 9-12.
2. Click Freeze Frame once more or press F6 again to let the ECG
continue.
Save 10s or F5 Saves the last 10 seconds of an ECG strip. The main measurement
values for the strip are indicated.
The system retrospectively saves the ECG strip which ends
immediately with the mouse click. You can mark this ECG strip.
Note: Changing the default speed to 25 mm/second or faster, will
not allow a full 10 second strip to appear in the window. If this
situation occurs, the ECG trace may show a temporary overlap after
Save 10s is clicked. This effect lasts only until the next sweep of the
display. Stored ECGs are not affected.
See Editing the analysis for this ECG strip on page 9-8.
Start ECG strip Saves an ECG strip of maximum 30 seconds. You can mark or delete
this ECG strip.
1. Click Start ECG strips.
The system starts saving the continuous ECG as an ECG strip.
2. During the saving click Stop ECG strip to stop the strip
prematurely.
The main measurement values for the last 10 seconds of the strip
are indicated.
Rhythm Enables continuous printing. When you click the button to start
continuous printing, the button changes color and prompts Click to
stop.
Note: You must manually stop continuous printing.
To print:
1. Click Rhythm.
The rhythm queue prints channels,
depending on the print
parameters. It can take several minutes until a complete page is
printed.
2. To stop printing, click Printing (Click to Stop).
Control
Panel Option Description