Product Manual

70-00533-02 A CareCenter MD 1.0
4-2
Acquire a Resting ECG
Operating steps - overview
The system uses the orange (main) button (or the F2 key) to perform a typical workflow.
Use the orange button (or F2) to start and to save an automatic 10 seconds 12-channel
ECG acquisition. Shaded steps are optional. For more information on each step see
Detailed procedures on page 4-4.
Step Selection
1
Search for a patient or enter a new patient. Patient Search tab or Patient...
2
Select the test procedure. Resting 12 or Ctrl + <number>
Options
If multiple acquisition devices are installed, select the device. Continue or press F2
If using a Bluetooth wireless device, power on the device.
3
Apply the electrodes:
Incorrectly applied electrodes blink.
The graphs indicate the hook up quality, check the ECG traces.
Options
Enable the filter only if you cannot achieve sufficient
acquisition quality.
EMG filter
Enter a remark. Remarks or press F3
Start a blood pressure measurement device or enter a blood
pressure value manually.
Blood pressure icon or press
Shift + F3
4
Start the acquisition.
Note: The button is active only if all electrodes are applied correctly.
Start Auto or press F2
Options
Start an acquisition manually if the electrodes are not applied
correctly.
Start Auto or press Ctrl + A
The system starts a timer which stops the record after 10
seconds. If you restart the timer within 10 seconds, the
acquisition starts from the beginning.
Reset or press Ctrl + R
Start and stop the acquisition manually. Start or press Ctrl + S
Stop or press Ctrl + T
5
Review the recording.
Options
Repeat the acquisition in manual or in automatic mode. Repeat or press Ctrl + S
Repeat Auto or press Ctrl + A
Confirm the diagnosis.
The system adds your full name and the date and time to the
diagnosis or to the remark.
Confirm or press Ctrl + B
Preview using the standard formats. Preview or press Ctrl + F
Print the recording in the standard format. Print or Ctrl + P