Product Manual
CareCenter MD 1.0 70-00533-02 A
3-25
Pacemaker detection
Pacemaker detection
The settings for pacemaker detection can be changed during the ECG acquisition.
To change pacemaker detection:
1. Open Settings | Pacemaker detection...
2. Select the electrodes to be used for the detection.
3. Select whether the pacemaker should be replaced with 1 mV impulse.
Pacemaker impulses
Impulses delivered by pacemakers can be so narrow that it is not possible to display the
original pulse. Therefore, the system shows a synthetic impulse exactly at the point of the
ECG where the original impulse was detected.
The synthetic impulse is displayed on all channels and always with positive polarity.
Figure 3-5: Example of pacemaker impulses
The level of the synthetic impulse is 1 mV. By default, the width of the synthetic impulse
is set to 8 ms.
Configuration Options
Change the value of the
synthetic impulse
1. Select File | Options... | Devices, and open CareCenter MD
(BT) or CareCenter MD (USB).
2. Enter the desired width in the Pacemaker Impulse Width
(ms) input box. The value must be between 2 ms and 16 ms.