ECG Acquisition
Pediatric ECG
For patients 16 years and under, the program does not create diagnostics. Instead, the
program creates a measurement table. In this table, the measurement values of the patient
are compared with those from statistics distributed after Gauss. The tables use the same
naming convention as the tables created by Andrè Davignon et al.
If the values of the patient are beyond this distribution, these are marked with asterisk (*).
Configuration Options
Configure the pediatric ECG
statistical distribution
1. Select File | Options… | 12 Lead ECG | ECG analysis.
2. In Table for pediatric ECG, select the gauss distribution for
the table:
– None
– 2% to 98%
– 5% to 95%
Values beyond this distribution are marked with an asterisk
(*)
Configure a default age and
gender to use for analysis
• Select File | Options … | 12 Lead ECG | ECG analysis.
– To set the default age, select If DoB is missing, analyze
with an age of and select the age.
– To set the default gender, select If gender is missing,
analyze with and select the gender.