CareCenter MD 1.0 70-00533-02 A
3-23
Change the ECG representation
Toggle views
You can cycle through all views which consist of only one column and have the same
number of leads.
Click Refresh to toggle the views.
If you selected a 3-channel or a 6-channel representation, you can use the refresh in the
header to switch to the other lead groups.
Change the mode of representation
The mode of representation affects only multi-column representations.
Synchronous modethe waveforms in the left column and the waveforms in the
right column are written simultaneously.
Sequential modethe waveforms in the left column and the waveforms in the right
column are written consecutively.
Advantagethe waveform is overwritten only half as often. This allows longer
viewing of the ECG.
Disadvantage—the ECG within a column is updated only half as often.
Configuration Options
The synchronous mode is the default.
Enable the sequential mode for
each procedure and for every
view
1. Select File | Options… | 12 Lead ECG | Test procedures.
2. Select the Display tab.
In all representations which are displaying real-time ECG, you
can select a multi-column view. To activate these views select
the Sequential mode option.