Product Manual

70-00533-02 A CareCenter MD 1.0
3-18
ECG Acquisition
You can change the filter settings for an individual test:
Default filter settings
Table 3-4 lists the default filter settings.
To... Do this...
Enable the Mains (AC line) filter Select Settings | Mains (AC Line) filter.
Enable the EMG (muscle
Artifact) filter
Click EMG filter or select Settings | EMG (Muscle Artifact)
filter.
Enable the filter defaults
(including the Baseline filter)
1. Select File | Options… | 12 Lead ECG | Test procedures.
2. Select the ECG tab and configure the settings.
Table 3-4: Default filter settings
Filter STAT ECG Resting ECG Rhythm ECG Stress Test
Mains (AC Line) filter On On Off On
EMG (Muscle Artifact) filter On Off Off On
Baseline filter On On On On