Product Manual

70-00533-02 A CareCenter MD 1.0
3-4
ECG Acquisition
The system includes further investigation programs which are not set.
Table 3-2: Additional test types
Test type Settings
Resting 6 Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Resting ECG
I II III aVR aVL aVF
Automatic recording stop after 10 seconds
No interpretation if less than 12 standard leads
Nehb Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Resting ECG
I II III D A J
Automatic recording stop after 10 seconds
No interpretation if less than 12 standard leads
Frank Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Resting ECG
I II III X Y Z
Automatic recording stop after 10 seconds
No interpretation if less than 12 standard leads
Rhythm 6 Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Rhythm ECG
I II III aVR aVL aVF
Automatic recording stop after 30 seconds
No interpretation if less than 12 standard leads
Resting Dorsal Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Resting ECG
I II III aVR aVL aVF V4 V5 V6 V7 V8 V9
Automatic recording stop after 10 seconds
No interpretation if less than 12 standard leads
Resting right
precordial
Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Resting ECG
I II III aVR aVL aVF V1 V2 V3R V4R V5R V6R
Automatic recording stop after 10 seconds
No interpretation if less than 12 standard leads
Rhythm Dorsal Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Rhythm ECG
I II III aVR aVL aVF V4 V5 V6 V7 V8 V9
Automatic recording stop after 30 seconds
No interpretation if less than 12 standard leads
Rhythm right
precordial
Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Rhythm ECG
I II III aVR aVL aVF V1 V2 V3R V4R V5R V6R
Automatic recording stop after 30 seconds
No interpretation if less than 12 standard leads