Product Manual

CareCenter MD 1.0 70-00533-02 A
3-3
Test procedure programs
Treadmill Test
Modified Bruce
Type of acquisition:
Leads:
Duration:
Arrhythmia monitoring:
Stress protocol:
Stress Test
I II III aVR aVL aVF V1 V2 V3 V4 V5 V6
The complete test is acquired
yes
Modified Bruce:
Load start with 1.7 mph, 0% grade
Increase by up to .9 mph, up to 5% grade
Stage duration 3 minutes
Stores a 10-second ECG strip at the end of
every stage
Blood pressure measurement at the end of
every stage
Pharmacological Type of acquisition:
Leads:
Duration:
Arrhythmia monitoring:
Stress protocol:
Stress Test
I II III aVR aVL aVF V1 V2 V3 V4 V5 V6
The complete test is acquired
yes
Pharmacological Stress Test:
No exercise device
Stage duration varies
Bicycle Stress Test
Standard
Type of acquisition:
Leads:
Duration:
Arrhythmia monitoring:
Stress protocol:
Stress Test
I II III aVR aVL aVF V1 V2 V3 V4 V5 V6
The complete test is acquired
yes
WHO:
Load start with 25 Watts
Increase by 25 Watts
Stage duration 2 minutes
Stores a 10-seconds ECG strip at the end of
every stage
Blood pressure measurement at the end of
every stage
Table 3-1: Basic test types (continued)
Test type Settings