Product Manual

70-00533-02 A CareCenter MD 1.0
3-2
ECG Acquisition
Test procedure programs
During the installation, the test type programs are set up. You may modify these programs
or create new ones.
The acquisition device unlocks specific test procedures. You can only complete recordings
for which you have the appropriate permissions.
Observe the instructions for the type of acquisition and options, as listed below. For more
information see the System Setup Guide.
Table 3-1: Basic test types
Test type Settings
STAT ECG Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Resting ECG
I II III aVR aVL aVF V1 V2 V3 V4 V5 V6
Records ECG until stopped manually
yes
STAT Rhythm ECG Type of acquisition
Leads
Duration
ECG Interpretation
Rhythm ECG
I II III aVR aVL aVF V1 V2 V3 V4 V5 V6
ECG recording until stopped manually
yes
Resting 12 Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Resting ECG
I II III aVR aVL aVF V1 V2 V3 V4 V5 V6
Automatic recording stop after 10 seconds
yes
Rhythm 12 Type of acquisition:
Leads:
Duration:
ECG Interpretation:
Rhythm ECG
I II III aVR aVL aVF V1 V2 V3 V4 V5 V6
Automatic recording stop after 30 seconds
yes
Treadmill Test
Bruce
Type of acquisition:
Leads:
Duration:
Arrhythmia monitoring:
Stress protocol:
Stress Test
I II III aVR aVL aVF V1 V2 V3 V4 V5 V6
The complete test is acquired
yes
Bruce:
Load start with 1.7 mph, 10% grade
Increase by up to .9 mph, 2% grade
Stage duration 3 minutes
Stores a 10-second ECG strip at the end of
every stage
Blood pressure measurement at the end of
every stage