Product Manual
CareCenter MD 1.0 70-00533-02 A
9-7
Displaying a Rhythm ECG record
Analysis Results
To obtain interpreted results, the system requires a test that includes at least the leads: I, II,
and V1 to V6. For a measurement, the acquisition must include leads I and II. The
Rhythm 12 test procedure fulfils these conditions for interpretation.
The system also requires a recording with at least 10 seconds of disturbance-free ECG and
patient information with at least age and gender. Age and gender may be default values.
If the system cannot analyze the data, the system displays the Summary only.
The Analysis section displays these measurement values and interpretations:
◆ Heart Rate
◆ Main measurement values
◆ Angle and amplitude of the waves as well as position type
◆ Indexes
◆ QT dispersion
◆ Rhythm line and diagnosis statements
Note: See Rhythm line on page 9-15 for descriptions of the characters used within the
Rhythm diagram. See the Hannover ECG System HES© Physician’s Guide for more
information about the ECG analysis program.
Configuration Options
Configure the display • Select: File | Options… | 12 Lead ECG | ECG analysis and
configure:
– Indication of the results of analysis in a different language
– Indication of the position types
– A simple or detailed diagram of the rhythm line
– Indication of the indexes
– Indication of the interference voltage
– Indication of the QT dispersion for all leads or for chest wall
leads
– Indication of the table with pediatric ECG