Displaying a Rhythm ECG record
Indication bar
You can customize and save the indication of main measurement values and frontal vectors
in the header. The indication of heart rate, blood pressure values, main measurement
values and frontal vectors apply to an (analyzed) section.
To customize the indication bar from the view:
1. Right-click the indication bar and select Customize....
The system displays the customize dialog box.
2. Display or hide a value in the header.
Select the check box to display the value.
Clear the check box to hide the value.
Selection Example Description
HR Heart rate always refers to the ECG strip. The
selected tab indicates the time of the ECG strip.
For analyzed records, the system always
calculates the heart rate by averaging over 10
seconds.
BP The blood pressure value applies to the selected
ECG strip.
Main measurement values Includes all of the main measurements.
The main measurement values (for analyzed ECG
strips) of the averaged beat display.
These values are also used in the Analysis
Results panel and in connection with the
indication of the enlarged averaged beat.
P P-duration
PR PR-interval
QRS QRS-duration
QT QT-interval
QTc Corrected QT-interval
Where: QTc = QT/
RR