70-00533-02 A CareCenter MD 1.0
7-2
Emergency ECG
Settings
Use these settings to configure the emergency ECG. While you can change these settings
at any time, they are usually configured just once after the installation.
Recording type
The system is configured for both the STAT ECG (Resting Emergency ECG) and the
STAT Rhythm ECG test procedure. The Rhythm test procedure allows you to record the
ECG for a long time. During the recording or later you can select multiple ECG strips for
analysis or interpretation.
Start option
You can configure the system to always start with an emergency test procedure.
Age and gender
The automatic analysis program requires the age and gender of the patient in order to
calculate measurement values and perform the interpretation. Because this data may not
be available for the emergency ECG, the recording includes no analysis unless you
configure a default age and gender.
Configuration Options
Allow only a single emergency
procedure
• Select File | Options... | 12 Lead ECG | Test procedures clear
the STAT ECG or STAT Rhythm ECG option.
Configuration Options
Change the start up default to
the emergency test procedure
1. Select File | Options... | General | Environment.
2. Select Startup with test procedure.
3. Select the emergency test procedure. (STAT ECG or STAT
Rhythm ECG).
Configuration Options
Set a default age and gender 1. Select File | Options... | 12 Lead ECG | ECG analysis.
2. Select age and gender for use in the analysis program.
Note: The output of Analysis Results uses the default
gender and age for diagnosis.