3.1 General Maintenance Information
is section contains information for ordinary upkeep of the BUFFALO FILTER ViroVac™ Smoke
Evacuation System. While the system has been designed and manufactured to high industry
standards, it is recommended that periodic inspection and performance testing be performed by
a qualied Biomedical Technician to ensure continued safe and eective operation.
3.2 Cleaning
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water.
Wipe dry with a clean cloth. Do not steam sterilize.
3.3 Periodic Inspection
e ViroVac™ Smoke Evacuation System should be visually inspected at least every year. is
inspection should include checks for:
• Damage to the power cord.
• Damage to the power plug or power inlet module.
• Proper mating, cleanliness and absence of damage to the lter inlet.
• Obvious external or internal damage to the system.
3.4 Troubleshooting the System – see below.
Maintenance
Section 3.0
PROBLEM POTENTIAL CAUSE CORRECTIVE ACTION
1. Smoke Evacuation System
is ON but suction is minimal
or none.
2. Smoke Evacuation System
does not function even
though suction ON button is
depressed.
1. Filter is not seated
completely.
2. Filter is clogged.
3. Vacuum hose or tube is
clogged.
4. Motor/blower is obstructed.
1. Not plugged into an
electrical outlet.
2. Fuses are blown.
3. Electronic system failure.
4. Filter life has expired or
invalid lter inserted.
1. Re-install ViroSafe Filter, press rmly into place
and fully engage clip.
2. Replace lter with a genuine Bualo Filter
ViroSafe lter.
3. Replace vacuum hose or tube with genuine
Bualo Filter products.
4. Call BioMed or BUFFALO FILTER Technical
Services at 1.800.343.2324 or 716.835.7000.
1. Check power outlet and connection to rear or
side panel of the machine.
2./3. Call BioMed or BUFFALO FILTER Technical
Services at 1.800.343.2324 or 716.835.7000.
4. Replace lter with a genuine Bualo Filter
ViroSafe lter.