Manual

CAUTION: Insure uid trap is
fully seated into the lter prior to
capturing uid.
4. Remove the appropriate cap on
the front of the uid trap based on
the size of the tubing that will be
used in the surgical application.
Store the cap on the appropriate
cap holder on top of uid trap.
5. Insert the chosen tubing
accessory into the appropriate
port located on the uid trap.
6. The uid trap is equipped with
measurement increments located
on the front of the device to
measure the amount of uid
captured.
WARNING: Insure that the uid
trap is not over-lled. Once the
uid level reaches 100cc, the
uid trap should be changed.
7. Once lled, or at the end of
each procedure, remove the
tubing accessory and place the
appropriate cap stored on the
top of the uid trap onto the
open port.