Manual
CAUTION: Insure uid trap is
fully seated into the lter prior to
capturing uid.
4. Remove the appropriate cap on
the front of the uid trap based on
the size of the tubing that will be
used in the surgical application.
Store the cap on the appropriate
cap holder on top of uid trap.
5. Insert the chosen tubing
accessory into the appropriate
port located on the uid trap.
6. The uid trap is equipped with
measurement increments located
on the front of the device to
measure the amount of uid
captured.
WARNING: Insure that the uid
trap is not over-lled. Once the
uid level reaches 100cc, the
uid trap should be changed.
7. Once lled, or at the end of
each procedure, remove the
tubing accessory and place the
appropriate cap stored on the
top of the uid trap onto the
open port.