Instruction Manual
Bualo Filter Turbo Auto-Sense™ Operator’s Manual: Page 11
TURBO MODE
e TURBO button is located on the key pad (Figure 1). e amount of suction is signicantly
increased when this button is pressed. To return to NORMAL mode, press the TURBO button
again.
FILTER LIFE REMAINING
e exact number of hours and minutes remaining on the ViroSafe® lter (in use) are displayed
on the LCD screen. When the lter’s life is consumed, time remaining will indicate that there are
0 hours and 0 minutes remaining. is line will continually ash. When the lter life reaches this
point, a new lter with time remaining should be inserted; however, the suction will still work.
When the unit’s main power is turned o or the system is le idle for more than 6 hours, a new
lter or a lter with remaining lter life MUST be inserted into the lter basket to allow the unit’s
suction to turn back on.
When the lter’s life is exhausted the
LCD screen will ash “REPLACE FILTER.”
FOOTSWITCH
A pneumatic footswitch is provided. To activate the footswitch, simply plug the footswitch tubing
into the jack ( ) on the rear panel. e suction can now be switched ON/OFF by depressing the
footswitch.
EZ LINK™ REMOTE ACTIVATION DEVICE (EZLINK01)
For directions on using the EZ Link™, see instructions accompanying the EXLINK01.
CIRCUIT FUSES
Two 10 amp circuit fuses (8 amp for International systems) are located on the rear panel within
the power entry module. ese circuit fuses protect both the system and the operator from dam-
age or injury.
SERVICE
When a service message appears, please contact Bualo Filter Customer Service for instructions.
Below are some examples of service screens that may appear.