Manual

BÜCHILabortechnikAG Installation|5
UltrasonicController 27/46
5.4 Feedtube,selection
1. Chooseafeedingtubefromthetablebelow.
NOTE
Forshortprocesses,evenunsuitabletubescanbeused,asswellingtakes
sometime.
Solvent Siliconetube Tygon
®
MH2075 Tygon
®
F4040A
Methanol
Ethanol
Acetone
Toluene
Isopropanol
Chloroform
Dichlormethane
Tetrahydrofuran
Ethylacetate
Hexane
Table1:=suitable|=unsuitable
5.5 Liquidfeedrate,adjustment
1. Adjustthefeedrateaccordingtothefollowingsettingsaccordingtothe
desiredrequirement:
Possibleliquidfeedrateusingultrasonicnozzle1–9mL/min(3–30%).
Recommendedfeedratewithaqueousfeeds2.5–3.2mL/min(8–10%).
5.6 Coolinggas,adjustment
Adjusttheflowrateofthecoolinggasusingtheneedlevalveattheflowmeter
oftheB290.
Thetemperatureoftheultrasonicnozzleshallalwaysbebelow100°C.Ifthe
temperatureoftheultrasonicnozzlerises(thetemperaturecanbecheckedon
theupperdisplayoftheultrasoniccontroller),theflowrateofthecoolinggas
needstobeincreased.