Owner manual

7 Maintenance
63 SpeedExtractor E-916/E-914 Operation Manual, Version D
7 Maintenance
This chapter provides instructions on all required maintenance to keep the instrument in good working
condition.
!
WARNING
Death or serious burns by electric current at cleaning.
Switch off the instrument
Disconnect the power cord and prevent unintentional restart
Wait until the instrument is completely dry before reconnecting the power supply
NOTICE
Risk of housing and instrument damage by liquids and detergents.
Do not spill liquids over the instrument or parts of it
Wipe off any liquids instantly
Use ethanol or soapy water as detergent only
7.1 Daily maintenance
Daily maintenance can prolong the system lifetime, reduce costs of service and downtime.
Fill the solvent reservoirs according to section 6.2.1. Make sure that the filter is always completely
dipped into the solvent.
Empty the waste bottle, if needed.
Inspect the cup seals for visual damage or contamination by sand or dust. If necessary, replace
the seals according to section 7.2.1. Always perform a leak test after replacing the seals (see sec-
tion 6.2.5).
Check the nitrogen pressure (6 10 bar).
Check the septa of the collection vials.
Inspect the needles for septa residue and/or deformation.
Cleaning the seals
If necessary, rinse the seals with organic solvent (e.g. ethanol) with the help of beaker. Always
visually inspect the seals for dust, sand or scratches.
NOTE
Never clean the cup seals with a wet towel or wipe. Any kind of residue might cause a leak and/or
reduce the lifetime of the cup seals.
Cleaning the upper cup seal frit
In case of unexpected contamination, remove the top cover plate (position 4 on page 97) and
clean it in an ultrasound bath.
Flushing the lines
If the next extraction method involves a different type of solvent, flush the lines thoroughly with the
new solvent as described in section 6.2.6.